>> back  
  Instructions for Use
 
[Product Name] Protective face mask                                       ■Non-sterile
[Model] N95-FWJ
[Construction] The mask consists of mask body, nose bar and ear loop. The mask body is a four-layer design, including a surface layer, two middle layers and a base layer, with the surface layer and base layer made of spun-bonded non-woven fabric, and the two middle layers made of polypropylene filter materials; the nose bar is made of a flexible plastic material that can be bent; the ear loop is made of nylon and spandex. The product is provided in non-sterile form. This respirator contains no components made from natural rubber latex.
[Application Method]
Step1: Open the package and take out the mask. Pic 1
Step2: Pinch it on the outsides with ear loops to unfold and then set it over your mouth and nose. Pic 2
Step3: For enhanced fit of the mask, tie the two ends of the ear loops to the back of the head with an attached clasp. Pic 3
Step4: Fit the mask and mold the nose bar around your nose bridge, to make sure the fit is tight and it snug around your face.
Pic 4~5
[Fit Testing] 
Cover the mask with both hands. Exhale sharply, if you feel gas leaking from the nose bridge, press the nose bridge firmly.If air leaks from the edge of the mask, readjust the wearing position to ensure tightness. The fit is critical to the performance of the product.
[Intend Use] The mask provides a physical barrier to filter particles in the air, droplets, blood, body fluids and secretions.
[Contraindication] Persons allergic to materials are prohibited to use.
[Transportation and storage] Transportation vehicle shall be clean and isolated from fire sources. The product should be stored in a dry, ventilated, non-corrosive gas environment with normal temperature, away from fire sources and flammable materials.
[Precautions]
Make sure the package is intact before use and check the mark, manufacture date and validity on the outer packing to ensure the product in validity.
Make sure it seals against your face.
Remove it within the same day you wear it. Once moisturized by sprays or sweater, replace it.
Check the fit every time you enter your working zone.
The product is provided in non-sterile form.
[Description] 1. Period of validity: 2 years
                    2. Explanation of graphics, symbols, abbreviations, etc. used in labels:


1 

3 

 4

2 

Consult instructions for use

Do not reuse

Don’t use when packing damaged

Non-sterile


[Manufacturer] Registrant / manufacturer /after-sales service: Qingzhou Yaowang Pharmaceutical Co., Ltd.
Residence of Registrant / Manufacturer: No. 1 Yaowangshan West Road, Qingzhou City, Shandong Province, China
Manufacturer address: No.3787, New york Road, Economy Development Zone, Qingzhou City, Shandong Province, China
Tel: +86-536-3261305 3261319 Zip code: 262515
Instructions for Use can be accessed at: http://www.qzywzy.com/n95.html
Medical Device Production License No.: Lu Shi Yao Jian Xie Sheng Chan Xu No.20200020
Medical Device Registration Certificate No. / Product Technical Requirements No.: Lu Xie Zhu Zhun 20202140285
Report Adverse Events, please send to yaowang@qzywzy.com.
[Executive standard] GB 19083-2010
[Issue date] June 19, 2020
Lot number, date of manufacture and expiration date are shown in the package.